Endress+Hauser Thermal Mass Flowmeters
Provides direct mass flow measurement for gases and selected liquid applications without the need for pressure or temperature compensation. Based on thermal dispersion technology, these instruments are designed for reliable monitoring of industrial gases, compressed air, nitrogen, biogas, and process air systems. The portfolio includes inline and insertion versions suitable for small pipelines up to large ducts.
Product Overview
The thermal mass portfolio includes:
• Proline t-mass F 300 / F 500 – Inline flanged versions
• Proline t-mass I 300 / I 500 – Insertion versions for large pipes and ducts up to DN 1500 (60″)
• Proline t-mass T 150 – Compact thermal flowmeter for liquids
• Flowphant T DTT31 / DTT35 – Flow switches for monitoring flow and temperature
These devices provide stable long-term performance and are designed for industrial process environments requiring reliable mass flow data.
Key Technical Features
• Direct mass flow measurement for gases
• No need for external pressure or temperature compensation
• Inline and insertion configurations
• Suitable for large pipe diameters (insertion versions)
• Reverse flow detection (model dependent)
• Unidirectional or bidirectional sensing options
• Compact or remote transmitter versions (up to 4 I/Os in 500 series)
• Robust stainless steel construction
• Integrated diagnostics and digital communication
Technical Specifications
Proline t-mass F 300 / F 500 (Inline – Flanged)
• Max measurement error (gas):
– 1.0% of reading (10–100% of full scale)
– 0.1% of full scale (1–10% of full scale)
• Measuring range: 0.5 to 3750 kg/h
• Medium temperature: –40 °C to +180 °C
• Max process pressure: PN40 / Class 300 / 20K
• Nominal sizes: DN 15 to DN 100
• Wetted materials: 316/316L stainless steel, CF3M, Alloy C22 options
Proline t-mass I 300 / I 500 (Insertion)
• Max measurement error (gas):
– 1.0% of reading (10–100% of full scale)
– 0.1% of full scale (1–10% of full scale)
• Measuring range: 20 to 733,501 kg/h
• Medium temperature: –40 °C to +180 °C
• Max process pressure: –0.5 to 20 bar(g)
• Suitable for pipes/ducts up to DN 1500 (60″)
• Wetted materials: 316/316L stainless steel, Alloy C22 options
• Clamping rings: PEEK, PVDF, or stainless steel
Proline t-mass T 150 (Liquid Monitoring)
• Flow accuracy: ±5% of full scale
• Measuring range: 226 to 14,100,000 l/h (reference conditions)
• Medium temperature: –20 °C to +100 °C
• Max process pressure: PN40
• Wetted materials: 316/316L stainless steel, Alloy C22 options
Standards & Compliance
• Industrial EMC compliance
• Hazardous area versions available (model dependent)
• Suitable for industrial gas and utility monitoring
• Designed for energy management and process optimization
Applications
• Compressed air monitoring
• Nitrogen and oxygen flow measurement
• Biogas and landfill gas
• Burner air control
• Utility gas monitoring
• Process gas measurement in chemical and industrial plants
• Large duct and ventilation systems (insertion versions)
Integration & System Compatibility
Thermal mass flowmeters integrate with PLC, DCS, and SCADA systems via analog and digital communication interfaces. The 500 transmitter platform supports remote installation and expanded I/O capability for advanced system integration.
Why Thermal Mass Flow Measurement
Thermal mass technology directly measures gas mass flow without requiring additional pressure or temperature inputs. This simplifies installation and reduces system complexity. The insertion versions allow cost-effective measurement in large pipelines without full-bore meter replacement.
TSB Local Support – Jordan
TSB provides application review, sizing support, system integration, and commissioning services for thermal mass flowmeters in Jordan. We support industrial facilities, utilities, and EPC contractors with correct selection and installation guidance.
Contact TSB Jordan for technical consultation and commercial support for Endress+Hauser thermal mass flowmeters.